Increasing informativity of digital radiographic systems
In this paper questions of improving the information content of radiographic images is considered. Source of the new information is X-ray images of inspected object registered in several energy ranges. Three energy X-ray images of test objects with an effective atomic number from 7.08 to 8.07 were o...
Gespeichert in:
Datum: | 2013 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | English |
Veröffentlicht: |
НТК «Інститут монокристалів» НАН України
2013
|
Schriftenreihe: | Functional Materials |
Schlagworte: | |
Online Zugang: | http://dspace.nbuv.gov.ua/handle/123456789/120107 |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Назва журналу: | Digital Library of Periodicals of National Academy of Sciences of Ukraine |
Zitieren: | Increasing informativity of digital radiographic systems / O.D. Opolonin, V.D. Ryzhikov // Functional Materials. — 2013. — Т. 20, № 4. — С. 528-533. — Бібліогр.: 9 назв. — англ. |
Institution
Digital Library of Periodicals of National Academy of Sciences of UkraineZusammenfassung: | In this paper questions of improving the information content of radiographic images is considered. Source of the new information is X-ray images of inspected object registered in several energy ranges. Three energy X-ray images of test objects with an effective atomic number from 7.08 to 8.07 were obtained. Using a photodiode-scintillator type detectors based on scintillator ZnSe(Te) the accuracy of substances separation by the effective atomic number ∼5 % was achieved. We proposed a method for reconstructing the energy dependence of the mass attenuation coefficient. |
---|