A new Krakow scanning nuclear microprobe: performance tests and early application experienc

A new scanning nuclear microprobe (MP) with a short-length probe forming system was designed, installed and tested at the 3MV Van de Graaff accelerator in Krakow. The MP resolution of 3.3mm was reached for a 2.4 MeV proton beam in the high-current mode (≥100pA). The MP facility provides a local, non...

Повний опис

Збережено в:
Бібліографічні деталі
Дата:2001
Автори: Lebed, S., Lekki, J., Paszkowski, M., Polak, W., Potempa, A., Stachura, Z., Styczen, J.
Формат: Стаття
Мова:English
Опубліковано: Національний науковий центр «Харківський фізико-технічний інститут» НАН України 2001
Назва видання:Вопросы атомной науки и техники
Онлайн доступ:http://dspace.nbuv.gov.ua/handle/123456789/79241
Теги: Додати тег
Немає тегів, Будьте першим, хто поставить тег для цього запису!
Назва журналу:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Цитувати:A new Krakow scanning nuclear microprobe: performance tests and early application experienc / S. Lebed, J. Lekki, M. Paszkowski, W. Polak, A. Potempa, Z. Stachura, J. Styczen // Вопросы атомной науки и техники. — 2001. — № 3. — С. 71-73. — Бібліогр.: 9 назв. — англ.

Репозитарії

Digital Library of Periodicals of National Academy of Sciences of Ukraine
id irk-123456789-79241
record_format dspace
spelling irk-123456789-792412015-03-31T03:01:58Z A new Krakow scanning nuclear microprobe: performance tests and early application experienc Lebed, S. Lekki, J. Paszkowski, M. Polak, W. Potempa, A. Stachura, Z. Styczen, J. A new scanning nuclear microprobe (MP) with a short-length probe forming system was designed, installed and tested at the 3MV Van de Graaff accelerator in Krakow. The MP resolution of 3.3mm was reached for a 2.4 MeV proton beam in the high-current mode (≥100pA). The MP facility provides a local, non-destructive, quantitative elemental microanalysis using a Proton Induced X-ray Emission (PIXE) technique. As example of possible applications an analysis of a geological sample containing monazite crystals investigated by PIXE method is presented. 2001 Article A new Krakow scanning nuclear microprobe: performance tests and early application experienc / S. Lebed, J. Lekki, M. Paszkowski, W. Polak, A. Potempa, Z. Stachura, J. Styczen // Вопросы атомной науки и техники. — 2001. — № 3. — С. 71-73. — Бібліогр.: 9 назв. — англ. 1562-6016 PACS numbers: 29.27.Eg, 32.30.Rj, 91.65.Dt, 91.70.-c http://dspace.nbuv.gov.ua/handle/123456789/79241 en Вопросы атомной науки и техники Національний науковий центр «Харківський фізико-технічний інститут» НАН України
institution Digital Library of Periodicals of National Academy of Sciences of Ukraine
collection DSpace DC
language English
description A new scanning nuclear microprobe (MP) with a short-length probe forming system was designed, installed and tested at the 3MV Van de Graaff accelerator in Krakow. The MP resolution of 3.3mm was reached for a 2.4 MeV proton beam in the high-current mode (≥100pA). The MP facility provides a local, non-destructive, quantitative elemental microanalysis using a Proton Induced X-ray Emission (PIXE) technique. As example of possible applications an analysis of a geological sample containing monazite crystals investigated by PIXE method is presented.
format Article
author Lebed, S.
Lekki, J.
Paszkowski, M.
Polak, W.
Potempa, A.
Stachura, Z.
Styczen, J.
spellingShingle Lebed, S.
Lekki, J.
Paszkowski, M.
Polak, W.
Potempa, A.
Stachura, Z.
Styczen, J.
A new Krakow scanning nuclear microprobe: performance tests and early application experienc
Вопросы атомной науки и техники
author_facet Lebed, S.
Lekki, J.
Paszkowski, M.
Polak, W.
Potempa, A.
Stachura, Z.
Styczen, J.
author_sort Lebed, S.
title A new Krakow scanning nuclear microprobe: performance tests and early application experienc
title_short A new Krakow scanning nuclear microprobe: performance tests and early application experienc
title_full A new Krakow scanning nuclear microprobe: performance tests and early application experienc
title_fullStr A new Krakow scanning nuclear microprobe: performance tests and early application experienc
title_full_unstemmed A new Krakow scanning nuclear microprobe: performance tests and early application experienc
title_sort new krakow scanning nuclear microprobe: performance tests and early application experienc
publisher Національний науковий центр «Харківський фізико-технічний інститут» НАН України
publishDate 2001
url http://dspace.nbuv.gov.ua/handle/123456789/79241
citation_txt A new Krakow scanning nuclear microprobe: performance tests and early application experienc / S. Lebed, J. Lekki, M. Paszkowski, W. Polak, A. Potempa, Z. Stachura, J. Styczen // Вопросы атомной науки и техники. — 2001. — № 3. — С. 71-73. — Бібліогр.: 9 назв. — англ.
series Вопросы атомной науки и техники
work_keys_str_mv AT lebeds anewkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT lekkij anewkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT paszkowskim anewkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT polakw anewkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT potempaa anewkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT stachuraz anewkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT styczenj anewkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT lebeds newkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT lekkij newkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT paszkowskim newkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT polakw newkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT potempaa newkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT stachuraz newkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
AT styczenj newkrakowscanningnuclearmicroprobeperformancetestsandearlyapplicationexperienc
first_indexed 2025-07-06T03:17:19Z
last_indexed 2025-07-06T03:17:19Z
_version_ 1836865909410496512
fulltext A NEW KRAKOW SCANNING NUCLEAR MICROPROBE: PERFORMANCE TESTS AND EARLY APPLICATION EXPERIENC S. Lebed1, J. Lekki2, M. Paszkowski3, W. Polak2, A. Potempa2, Z. Stachura2, J. Styczen2 1 Institute of Applied Physics (IAP), UR-44030 Sumy, Ukraine, salmp1@yahoo.com 2 Institute of Nuclear Physics (INP), PL-31342 Krakow, Poland, Janusz.Lekki@ifj.edu.pl 3 Institute of Geological Sciences, Polish Academy of Science, PL-31002, Krakow, Poland A new scanning nuclear microprobe (MP) with a short-length probe forming system was designed, installed and tested at the 3MV Van de Graaff accelerator in Krakow. The MP resolution of 3.3µm was reached for a 2.4 MeV proton beam in the high-current mode (≥100pA). The MP facility provides a local, non-destructive, quantitative ele- mental microanalysis using a Proton Induced X-ray Emission (PIXE) technique. As example of possible applica- tions an analysis of a geological sample containing monazite crystals investigated by PIXE method is presented. PACS numbers: 29.27.Eg, 32.30.Rj, 91.65.Dt, 91.70.-c 1 INTRODUCTION The new scanning nuclear microprobe (MP) with a short probe forming system (Fig.1) of 2.3m length was designed, constructed and installed in the Institute of Nuclear Physics (IFJ) in Krakow, Poland, during the pe- riod from 1995 to 1999 [1-4]. Fig. 1. Microbeam line of the Krakow scanning nuclear microprobe. The accelerating and beam transporting systems of the Krakow MP were improved significantly last years. The MP parameters are maintained stable over many hours, with a relatively high ion microbeam current (up to 3nA at ≤10µm beam spot size) permitting measure- ments to be performed within a reasonable time. With such a current the microprobe proton induced X-ray emission (micro-PIXE) technique can be used in many applications of the geological sciences. The micro-PIXE is now accepted in many laboratories throughout the world as a highly sensitive method for elemental analy- sis. The method was developed at the Krakow MP [4,5]. The subject of this paper is to demonstrate the Krakow MP performance and micro-PIXE technique ca- pabilities in geological research. 2. EXPERIMENT 2.1 Accelerator The Krakow MP is based on a Van de Graaff (HVEC type K-3000) accelerator. Performance of the set-up has been improved by updating the accelerator during the year 2000: a new high-brightness RF ion source and a new belt charge power supply have been installed. 2.2 Beam transport system The previous beam transport system of the Krakow MP was based on a quadrupole doublet of magnetic quadrupole lenses [3]. This was changed by installing an additional magnetic quadrupole doublet [4]. The dou- blet (Fig. 2) was manufactured in the IFJ mechanical workshop. Fig. 2. New magnetic quadrupole doublet. The new optimised beam transport system is based on a “divided Russian quadruplet” geometry providing a minimum contribution of the intrinsic and parasitic aberrations to the degradation of the beam brightness in the system. The system optics and quadrupole doublet design were developed by S.Lebed. These innovations and other above-mentioned VdG modernisation have laid the foundations of a significantly improved MP per- formance. 2.3 MP performance The Krakow MP resolution is tested by measuring the beam profile and beam spot size (full width at half- maximum (FWHM)) at the target [4]. For this purpose a copper electron microscope grid is used. Secondary electron profiles from the grid are measured. The pro- files from linear beam scanning over the grid are anal- ysed. FWHM reaches 3.3µm (Fig. 3) in the high current (PIXE) mode and less than 10µm at 3nA proton beam ВОПРОСЫ АТОМНОЙ НАУКИ И ТЕХНИКИ. 2001. №3. Серия: Ядерно-физические исследования (38), с. 71-73. 71 mailto:Janusz.Lekki@ifj.edu.pl current (IP). Detectors used are: a channeltron for sec- ondary electron imaging and a Si(Li) detector for PIXE measurements. Fig. 3. Calibration copper grid (400mesh/inch) im- aged by PIXE (2.4 MeV protons) in Kα and Kβ X-ray lines. Left: spot size is 3.3µm at IP =100pA. Right: spot size is 8µm at IP =3nA. 2.4 Data acquisition system A new data acquisition and evaluation system was installed [5]. The system collects detector events in the list (event–by–event) mode, storing the number of counts, energy pulse height and the current x and y co- ordinates of the beam. This data allows detailed off–line sample analysis. The MP scanning mode also makes possible a dynamic on–line two–dimensional (2D) anal- ysis of the elements (trace, minor and major) in the sam- ple by indicating the regions of interest (ROI). 3 RESULTS AND DISCUSSIONS The geological micro–PIXE investigations were car- ried out on a sample of monazite crystals mounted in epoxy glue and coated with carbon (Figs. 4-7). The sample was prepared by processing the uppermost car- boniferous sandstone from Kwaczawa village, Gródek gorge, Poland. Fig. 4. Light microscope image (600x400 µm2) of the sample. In order to improve the minimum detection limit of heavy trace elements, a 200µm Al filter was placed in front of the Si(Li) detector to absorb most of the intense low energy X-rays emitted from the matrix materials of the sample. The relative intensities of the characteristic lines of the U, Th and Pb elements evaluated in the present experiment (Figs. 5, 6) can be used to calculate the chemical age of the geological sample, applying the Montel method [8]. Fig. 5. Total and trace micro-PIXE elemental maps of a large monazite crystal. Measurement time is 15 minutes, scan area is 256x256 µm2. Images show fairly homogeneous distribution of Pb, U and Th in the sample. Fig. 6. Micro-PIXE spectra of the two monazite crystals. Measurement time is 70 minutes, scan area is 16x16µm2. Fig. 7. Total and trace (with ROI set to Zr-Kα line) PIXE elemental maps of the monazite samples. Im- age resolution is 64×64 pixels, scan area is 256x256 µm2. A small zirconium inclusion (~50x100 µm2) is clearly seen in the central part of the image. The micro-PIXE method delivers high detection sen- sitivity (1–10 ppm) for a broad range of geochemically important elements with Z>11. An effective depth of the analysis is 10–35µm. It allows the analysis of specific micro–inclusions (e.g. relics of mineralising fluids) in the minerals [9], containing important genetic informa- tion. In industrial geology, trace element distributions in minerals are used in prospecting for diamonds, platinum group minerals as well as in petrology, mining, ore gen- esis, etc. [7–9]. 72 4. CONCLUSIONS The spatial resolutions achieved by the Krakow MP are currently 3.3µm at beam current of 100pA and 8µm at beam current of 3nA for a 2.4MeV protons. The mi- cro-PIXE technique is developed at the Krakow MP. The average U-Th-Pb trace elemental data obtained will be used to calculate the chemical age of a geological sample. Most MP applications in geology are connected with the use of micro-PIXE in trace element geochemistry of rock forming minerals and their inclusions. The Krakow MP is well suited for these purposes. The results obtained on the Krakow MP show the micro-PIXE technique to be highly sensitive, local, fair- ly fast, non-destructive, quantitative elemental micro- analysis. The MP will be applied as an analytical tool in geology, biophysics, medicine, environmental and mate- rials sciences, etc. ACKNOWLEDGEMENTS The work has been supported by the State Commit- tee of Scientific Research (KBN) in Poland (contract No.1438/IA/620/95) and the Ministry of Education and Science in Ukraine (contract No. 2M/289-99). REFERENCES 1. S.Lebed, Z.Cioch, A.Rys, Z.Stachura, L.Zrodlows- ki, M.Cholewa, D.Jamieson, G.Legge // Ukrainskij Fizicheskij Zhurnal. 1999, v .44, N 8, p. 937. 2. S.Lebed // Nucl. Instrum. and Meth. 1999, B 152, p. 145. 3. S.Lebed // Nucl. Instrum. and Meth. 1999, B 155, p. 322. 4. S.Lebed, Z.Stachura, M.Cholewa, G.J.Legge, J.Lekki, S.Maranda, A.Potempa, C.Sarnecki, Z.Szklarz, J.Styczen, B.Sulkio–Cleff // Nucl. In- strum. and Meth. 2001, B 181 (1-4), in print. 5. J.Lekki, R.Hajduk, S.Lebed, A.Potempa, T.Pieprzy- ca, Z.Stachura, M.Zieblinski, J.Styczen, Report No.1856/AP (2000), Inst. of Nuclear Physics Krakow, www.ifj.edu.pl/reports/2000.html. 6. S.Lebed, T.Butz, J.Vogt, T.Reinert, D.Speemann, J.Heitmann, Z.Stachura, J.Lekki, A.Potempa, J.Sty- czen, B.Sulkio–Cleff // Nucl. Instr. and Meth. 2001, B 181 (1-4), in print. 7. C.G.Ryan // Nucl. Instrum. and Meth. 1995, B 104, p. 377. 8. R.Frei, V.M.Prozesky, W.Przybylowicz // Nucl. In- strum. and Meth. 1997, B 130, p. 676. 9. S.H.Sie // Nucl. Instrum. and Meth. 1997, B 130, p. 592. ВОПРОСЫ АТОМНОЙ НАУКИ И ТЕХНИКИ. 2001. №3. Серия: Ядерно-физические исследования (34), с. 73-73. 73